PERBANDINGAN EFEKTIVITAS STERILISASI INSTRUMEN BEDAH METODE OVEN ANTARA SUHU 1600 C /120 MENIT, SUHU 1800 C / 20 MENIT DAN METODE AUTOKLAF STEAM SEBAGAI BAKU EMAS DI RSU HAJI SURABAYA

DJUMADI PURNOMO, NIM011141014 (2013) PERBANDINGAN EFEKTIVITAS STERILISASI INSTRUMEN BEDAH METODE OVEN ANTARA SUHU 1600 C /120 MENIT, SUHU 1800 C / 20 MENIT DAN METODE AUTOKLAF STEAM SEBAGAI BAKU EMAS DI RSU HAJI SURABAYA. Thesis thesis, UNIVERSITAS AIRLANGGA.

[img]
Preview
Text (ABSTRAK)
gdlhub-gdl-s2-2014-purnomodju-29498-8.abstr-t.pdf

Download (225kB) | Preview
[img] Text (FULL TEXT)
gdlhub-gdl-s2-2014-purnomodju-29498-full text.pdf
Restricted to Registered users only

Download (2MB) | Request a copy
Official URL: http://lib.unair.ac.id

Abstract

The population in this study is the Surgical Instruments, data processing: after the data is collected, the data processing is done, Editing, Codimg, Entry, tablature, and Dry subsequently processed in the form of a table data analysis: Statistical analysis using a 95% significance level (α = 0.05), and using computerized program spss . Results of normality test data do not show a normal distribution, then the analysis is done with non-parametric approach, using a comparison test between the two groups of samples, so since there are 3 groups of the sample (temperature) then there are 3 combinations of comparison, the oven temperature of 1600C to 1800C oven, oven temperature of 1600C to 1210C steam autoclave and oven temperature of 1800C to 1210C steam autoclave. Statistical test used is the Mann Whitney test. 1600C temperature comparison with 1800 C temperature, Mann Whitney test results obtained p value of 0.001, when using α 0.05 then we can conclude p value (0.001> 0.05), meaning that the temperature of 1600C for 120 minutes with a temperature of 1800C for 20 minutes there is a significant difference. 1600C temperature comparison with 1210 Mann Whitney test obtained p value of 0.003, when using α 0.05 then we can conclude p value (0.003> 0.05), meaning that the temperature of 1600C for 120 minutes with a steam autoclave temperature of 1210C for 30 minutes there is a significant difference. 1800C temperature comparison with 1210 Mann Whitney test results obtained p value of 0.001, when using α 0.05 we can conclude p value (0.001 <0.05), meaning that the temperature of 1800C for 120 minutes with a steam autoclave temperature of 1210C for 30 minutes there is a significant difference Based on the statistical analysis of the number of colony growth at different temperatures between 1600C oven temperature for 120 min, oven temperature 1800C for 20 min with 1210 temperature steam autoclave for 30 minutes, there were significant differences at different temperatures between the same oven.

Item Type: Thesis (Thesis)
Additional Information: KKA KK TKD. 62/13 Pur p
Uncontrolled Keywords: Surgical instruments, sterilization oven, autoclave steam sterilizer, bacilus stearotermophillus
Subjects: R Medicine > RA Public aspects of medicine > RA1-1270 Public aspects of medicine > RA960-1000.5 Medical centers. Hospitals. Dispensaries. Clinics
R Medicine > RD Surgery
Z Bibliography. Library Science. Information Resources > ZA Information resources > ZA4450 Databases
Divisions: 09. Sekolah Pasca Sarjana > Ilmu Kedokteran Dasar
Creators:
CreatorsNIM
DJUMADI PURNOMO, NIM011141014NIM011141014
Contributors:
ContributionNameNIDN / NIDK
Thesis advisorAryati, Dr., dr., M.S., Sp.PK(K)UNSPECIFIED
Thesis advisorI G A A Putri Rejeki, dr., Sp.PK(K)UNSPECIFIED
Depositing User: Nn Duwi Prebriyuwati
Date Deposited: 2015
Last Modified: 31 Aug 2016 09:52
URI: http://repository.unair.ac.id/id/eprint/38498
Sosial Share:

Actions (login required)

View Item View Item