Imprinted-Zeolite-X-Based Sensor for Non-Enzymatic Detection of Blood Glucose by Potentiometry

Miratul Khasanah, .- and Alfa Akustia Widati,, .- and Usreg Sri Handajani, .- and Akhsin Mastura,, .- and Eka Yunicha Sari., .- (2022) Imprinted-Zeolite-X-Based Sensor for Non-Enzymatic Detection of Blood Glucose by Potentiometry. Chem Engineering, 6 (5). 01-04. ISSN E-ISSN:2305-7084

[img] Text (Fulltext)
C05. Fulltext.pdf

Download (9MB)
[img] Text (Penilaian dan Validasi)
C05. Penilaian dan Validasi.pdf

Download (949kB)
[img] Text (Similarity)
C05. Similarity.pdf

Download (3MB)
[img] Text (Bukti Korespondensi)
C05. Bukti Korespondensi.pdf

Download (5MB)
Official URL: https://www.mdpi.com/2305-7084/6/5/71

Abstract

The development of sensors based on imprinted zeolite X to detect blood glucose through potentiometry was performed. In this study, the sensor was made of a mixture of carbon paste and imprinted zeolite X. Zeolite X was synthesized using a sol–gel-hydrothermal method at a temperature of 100 °C with basic materials of NaAlO2, NaOH, TEOS, and distilled water. The characterization results of XRD showed the presence of specific peaks, which were confirmed with standard zeolite X. Imprinted zeolite X exhibited a 20 times greater adsorption capacity size, and an adsorption efficiency 3 times greater than that of zeolite X. This is thought to be due to the presence of a molecular template within it. The IZ–carbon paste electrode showed optimum performance due to a mass ratio of carbon, paraffin, and imprinted zeolite X of 12:7:1. The electrode performance was expressed by the Nernst factor value of 30 mV/decade, the measuring range of 10−4–10−2 M, the upper detection limit of 1.38 × 10−2 M, and the lower detection limit of 1.28 × 10−4 M, so this electrode can be used for glucose analysis with a normal concentration (70–110 mg/dL or equivalent to 3.8 × 10−3–6.1 × 10−3 M), as well as the glucose concentration of people with diabetes mellitus (>200 mg/dL or about 10−2 M). This electrode showed precision values of 97.14–99.02%, accuracy values of 98.65–99.39%, and electrode response times of 10–13 s. The electrodes showed high stability for more than 5 weeks with 141 uses. The electrodes also showed high selectivity for glucose in the matrix of uric acid, urea, NaCl, and KCl. Therefore, its use as an alternative electrode for routine glucose analysis in the medical field is recommended.

Item Type: Article
Uncontrolled Keywords: imprinted zeolite X; potentiometric sensor; blood glucose; medical
Subjects: Q Science
Q Science > Q Science (General)
Q Science > QD Chemistry
Q Science > QD Chemistry > QD71-142 Analytical chemistry
Divisions: 08. Fakultas Sains dan Teknologi > Kimia
Creators:
CreatorsNIM
Miratul Khasanah, .-UNSPECIFIED
Alfa Akustia Widati,, .-UNSPECIFIED
Usreg Sri Handajani, .-UNSPECIFIED
Akhsin Mastura,, .-UNSPECIFIED
Eka Yunicha Sari., .-UNSPECIFIED
Depositing User: Mr Vega Andi Budiman
Date Deposited: 09 May 2023 00:36
Last Modified: 03 Jun 2023 07:16
URI: http://repository.unair.ac.id/id/eprint/125976
Sosial Share:

Actions (login required)

View Item View Item