VALIDASI METODE SPEKTROFOTOMETRI UV DAN KLT - DENSITOMETRI UNTUK PENETAPAN KADAR KOLKISIN DALAM TABLET

RATNA BINGAR IRNESTIN, 050911177 (2013) VALIDASI METODE SPEKTROFOTOMETRI UV DAN KLT - DENSITOMETRI UNTUK PENETAPAN KADAR KOLKISIN DALAM TABLET. Skripsi thesis, UNIVERSITAS AIRLANGGA.

[img]
Preview
Text (ABSTRAK)
gdlhub-gdl-s1-2014-irnestinra-29297-7.abstr-k.pdf

Download (161kB) | Preview
[img] Text (FULLTEXT)
gdlhub-gdl-s1-2014-irnestinra-29297-1.FULLTEXT.pdf
Restricted to Registered users only

Download (817kB) | Request a copy
Official URL: http://lib.unair.ac.id

Abstract

A research has been conducted to compare UV spectrophotometry and thin layer chromatography (TLC)-densitometry method to determine colchicine in tablet. The objective of this research is to validate the methods to fulfill the requirements of method validation for UV spectrophotometry and TLC-Densitometry. Selectivity, linearity, accuracy, and precision are the measured validation parameter in this research. Chloroform : aquadest (30:20) are the immiscible solvent used for extracting colchicine from tablet matrices. Selectivity was measured in UV spectrophotometry at 350 nm to analyze colchicine. In TLC-densitometry the selectivity was measured by applying colchicine into silica gel 60 F 254 with chloroform : aceton : diethylamine ( 5 :4 :1 ) mixture as the mobile phase. The spot scanned at 353 nm and its Rf value was 0.55. Linearity analysis with UV spectrophotometry using colchicine standard solution 8 to 20 ppm (r=0.9996, Vxo= 0,67 % ) and analysis with TLC-densitometry using 200 to 1000 ppm (r=0.9984, Vxo= 6,54 %). Accuracy and precision analysis were done using 80%, 100%, and 120% of therapeutic concentration (500 μg). By UV spectophotometry, the average recovery was 90.83% with RSD 3,16%, while TLC-densitometry had average recovery 86.11% with RSD 6.48%. Accuracy in both method meet the requirement (average recovery = 85% - 110%) but the TLC-densitometry was not fulfill the precision requirement (RSD ≤ 4 %). However, the percentage of colchicine recovery using UV spectrophotometry compared to the concentration in the label is 102.5% ± 3,40 %. This is not exceed the percentage labeled amount claimed, ( not less than 90 % and not more than 110 % USP XXXII )

Item Type: Thesis (Skripsi)
Additional Information: KKB KK-2 FF.KF 37/13 Irn v
Uncontrolled Keywords: VALIDASI METODE
Subjects: Q Science > Q Science (General) > Q1-390 Science (General)
Q Science > QD Chemistry > QD1-999 Chemistry
Divisions: 05. Fakultas Farmasi
Creators:
CreatorsNIM
RATNA BINGAR IRNESTIN, 050911177UNSPECIFIED
Contributors:
ContributionNameNIDN / NIDK
Thesis advisorDr. Hj. Isnaeni, Apt., M.S., Dr. Hj., Apt., M.S.UNSPECIFIED
Depositing User: shiefti dyah alyusi
Date Deposited: 08 Jan 2014 12:00
Last Modified: 10 Aug 2016 07:23
URI: http://repository.unair.ac.id/id/eprint/9729